Complications of Motility Coupling Post - Porous Polyethylene Orbital Implant Coupling System

多孔性聚乙烯義眼經栓子植入後引起之併發症

Abstracts


目的:針對多孔性聚乙烯義眼經栓子植入後引起之併發症提出臨床報告。病人與方法:共有二十位接受多孔性聚乙烯義眼植入的病人,其中十二位之後再接受栓子植入,我們記錄並分析術後的併發症。結果:二十位病人接受多孔性聚乙烯義眼植入後,未發現明顯的併發症。多孔性聚乙義眼植入與栓子植入的間隔時間平均為七點二個月,十二位接受栓子植入的病人中,有四位(33.3%)發生併發症,包括兩位栓子掉入眼窩中,一位聚乙烯義眼暴露,一位栓子栓入角度偏斜。結論:多孔性聚乙烯義眼經栓子植入後有幾種可能的併發症,其中最嚴重的是栓子掉入眼窩中。必須有更長的追蹤時間與更多的病例來確定多孔性聚乙烯義眼經栓子植入後的併發症。

Parallel abstracts


Objective: We report our clinical experience with the Motility Coupling Post (MCP)-porous polyethylene orbital implant (PPOI) coupling system and its complications. Patients and Methods: In this retrospective study, we collected 20 patients who underwent implantation of PPOIs. Twelve patients went on to have an MCP placed. The complications were analyzed. Results: There were no significant complications with PPOIs. All sockets showed good motility. The mean time interval between the implantation and MCP placement was 7.2 months. 4 of 12 patients (33.3%) who underwent MCP placement were found to have problems, including loss of the MCP in the orbital cavity (2 of 12), implant exposure (1 of 12), and MCP drilled on an angle (1 of 12). Conclusions: Implanted MCPs have several potential complications. The most serious problem is loss of the MCP into the orbital cavity. A longer follow-up with a larger number of patients is required to determine the complications of the MCP-PPOI coupling system in humans.