With laser turbidimeters and LED turbidimeters developed in our laboratories, total SO_x concentrations of some stack gas samples have been determined. Total SO_x concentration of each sample is also determined with traditional chemical precipitation-titration method. The correlation between the results of these measurements is very satisfactory. Turbidimetry may be used in even lower concentration range.