摘要


為了解市售電子血壓計之品質,本計畫於民國八十年十月至八十一年六月間,就臺灣全省各地區抽購檢體共三十件,依行政院衛生署自量式電子血壓計標準規格草案,對電子血壓計壓力換能器/標示器之精確度進行測試。檢測結果,除一件檢體於測試時似已損壞,因而不符規定外,其他產品均合格,不合格率相當於3.33%,較民國七十七年品質調查報告所得產品不合格率12.87%低約十個百分點,顯示短期而言,本類產品品質並不因其改列為無須辦理查驗登記之醫療器材項目而受影響。

並列摘要


A total of 30 samples of commercially available electronic sphygmomanometers were randomly collected from throughout Taiwan between October 1991 and June 1992. Accuracy of the pressure transducer/indicator were investigated following official methods. The results show that only a single sample (3.33%) failed to meet the government regulation.