透過您的圖書館登入
IP:18.225.149.143
  • 期刊

市售疼痛舒緩用經皮神經電刺激器之輸出電性安全測試、功能控制鍵測試、短路與開路測試及電池極性反向測試品質調查

An Investigation on the Output Characteristics Tests、Functional Controls Tests、Open/Short Circuit Performance Tests、Incorrect Battery Installation Tests of Transcutaneous Electrical Nerve Stimulators

摘要


本計畫自民國九十一年五月至六月,由本局同仁自行至台北縣市之醫療器材行、保健器材行、百貨公司及藥局抽購「經皮神經電刺激器」檢體共計20件。經以美國AAMI NS4「經皮神經電刺激器」標準之方法、規格進行檢測,共包括輸出電性安全測試、功能控制鍵測試、短路與開路測試及電池極性反向測試等四個項目。測試結果:共3件(15.0%)不符合美國AAMI NS4之輸出電性安全測試規定;此外,本次抽購的20件檢體中,有5件未依規定辦理查驗登記。

並列摘要


In order to ascertain the quality of Transcutaneous Electrical Nerve Stimulators, 20 samples were purchased from medical supplyers, department stores and pharmacies from Taipei county/city during the period from May to June 2002. Four tests which included output characteristics tests, functional controls tests, open/short circuit performance tests and incorrect battery installation tests were conducted according to the methods listed in ANSI/AAMI NS4 for Transcutaneous Electrical Nerve Stimulators. The results showed that three samples (15.0%) failed to meet the requirements in output characteristics tests. In addition, among these 20 samples, five samples did not obtain pre-market approval from Department of Health (DOH).