透過您的圖書館登入
IP:18.221.174.248
  • 期刊

Exploration and Application of Ion Selective Electrode Method for the Determination of Fluoride in Soil

摘要


The water-soluble fluoride in soil samples was determined by ion-selective electrode method, and good results were obtained. The detection limit was 0.314 mg/kg and the lower limit was 1.26 mg/kg. The relative standard deviation (n=7) between 3.82% and 6.28%, and the relative error (n=7) between 1.6% and 5.8% were obtained by the precision and accuracy analysis of standard materials (GBW07412a, GBW07414a, GBW07459). The test results show that the relative standard deviation and relative error of the test sample are within the allowable range.

參考文獻


Jiang Qian, Han Yong, Sun Xiaoli, et al. Study on water-soluble fluorine extraction method in acid soil [J]. Soil, 2012, 44(1): 140-143.
Zhou Yi, Chen Xin, Li Cheng, et al. Study on the optimum conditions of water-soluble fluorine in soil by deionized water extraction [J]. Chinese Journal of Health Inspection, 2010, 20 (8): 2073-2074.
Zhang Wei, Du Hao, Zhang Yanyan, et al. Thoughts on improving the supervision and management of provincial soil environmental quality [J]. Environmental Monitoring and Early Warning, 2012, 4(3): 53-56.
Yao Jianting, Wang Hui, Wang Qian. Study on Determination of Fluoride in Soil by Ion Selective Electrode Method[J]. Anhui Agricultural Sciences, 2009, 37( 5) : 1899-1900.
Wang Wenbao, Wang Xiaochun, Teng Ying, et al. Evaluation and analysis of soil heavy metal pollution in the stainless steel industry cluster area of Jiangsu Province [J]. Environmental Monitoring and Early Warning, 2010, 2(5): 40-43.