USB驅動型迷你恆電位儀之開發與鄰苯二甲醛消毒劑之檢測

Development of a USB-powered Mini-potentiostat and the Detection of o-phthalaldehyde Disinfectant

指導教授 : 陳力騏
共同指導教授 : 陳仲瑄(Chung-Hsuan Chen)

摘要


本研究採用具有高輸入阻抗、低輸入偏壓電流與低輸入補償電壓等特性之運算放法器(OPA2336P,Texas Instruments),搭配USB供電的DAQ(NI 6008/6009,National Instruments)與LabVIEWTM撰寫的圖形化使用者介面,組裝一台平行輸入自動換檔並且便於攜帶之小型恆電位儀,可應用於實驗室外的現場檢測並且與商用實驗室泛用型儀器(AutolabTM,Eco Chemie)相比所得到之檢測結果幾乎相同。本研究所提出之恆電位儀操作電位的範圍約從-1.3 V到1.1 V,最佳量測的解析度可到100 pA。以本研究所提出之恆電位儀搭配流動式注射系統,對鄰苯二甲醛濃度做初步的的濃度檢測,所量測到的氧化電流大小在0.2%-0.65%(w/v)範圍內呈線性相關(R2 = 0.9911),且變異係數小於5%,靈敏度為0.18 μA/cm2 mM,可應用於評估消毒內視鏡常用之鄰苯二甲醛最低有效濃度(Minimum effective concentration, MEC)。

並列摘要


An auto ranging mini-potentiostat was constructed by parallel reading architecture and using operational amplifiers (OPA2336P, Texas Instruments) with high input impedance, low input bias current and low input offset voltage. After implementing with a USB-powered DAQ (NI 6008/6009, National Instruments) and LabVIEWTM based graphical user interface, the portable electrochemical device can be used for on-site analysis with a performance comparable to a commercialized laboratory machine (AutolabTM, Eco Chemie). The operating voltage ranges from -1.3 V to 1.1 V and the resolution can reach 100 pA level. The mini-potentiostat was served as the amperometric detector of a conventional flow injection analysis system for measuring o-phthalaldehyde concentration. The current outputs (the peak heights of FIAgram) are reproducible (CV < 5%) and showed good linearity (R2 = 0.9911) in the range of 0.2%-0.5% (w/v), which can be used to evaluate the antiseptic efficiency (minimum effective concentration of o-phthalaldehyde) of the disinfectant used for cleaning endoscope. The sensitivity is 0.18 μA/cm2 mM.

並列關鍵字

potentiostat o-phthalaldehyde disinfectant

參考文獻


3. 顏旭亨、陳淑慧、黃美麗、林麗珍。2012。內視鏡高層次消毒液之使用及管理。感染控制雜誌 22(4):163-169。
4. Ahmadi, M.M., G.A. Jullien. 2009. Current-mirror-based potentiostats for three-electrode amperometric electrochemical sensors. Circuits and Systems I: Regular Papers, IEEE Transactions on 56(7): 1339-1348.
5. Alfa, M.J., D.L. Sitter. 1994. In-hospital evaluation of orthophthalaldehyde as a high level disinfectant for flexible endoscopes. J. Hosp. Infect. 26(1): 15-26.
7. Avdikos, E.M., M.I. Prodromidis, C.E. Efstathiou. 2005. Construction and analytical applications of a palm-sized microcontroller-based amperometric analyzer. Sensors Actuators B: Chem. 107(1): 372-378.
8. Ayers, S., K.D. Gillis, M. Lindau, B.A. Minch. 2007. Design of a cmos potentiostat circuit for electrochemical detector arrays. IEEE Transactions on Circuits and Systems I: Regular Papers 54(4): 736-744.

延伸閱讀