Title

三葉片機械心瓣轉軸系統的改良與測試

Translated Titles

Modification and Testing of the Pivot System of a Trileaflet Mechanical Heart Valve

DOI

10.6846/TKU.2005.00012

Authors

陳子偉

Key Words

三葉片機械心瓣 ; 中央開孔 ; 體外模擬 ; 樞紐凹槽 ; trileaflet mechanical heart valve ; central flow ; mock circulate loop ; fillister

PublicationName

淡江大學機械與機電工程學系碩士班學位論文

Volume or Term/Year and Month of Publication

2005年

Academic Degree Category

碩士

Advisor

盧博堅

Content Language

繁體中文

Chinese Abstract

本研究所設計的三葉片機械心瓣,乃採用中央開孔式的結構,符合人體自然心瓣的運作方式。和傳統的單葉或雙葉機械心瓣相比,此三葉片機械心瓣的中央孔口,可加大中央孔口有效面積、避免血液分離、減少壓差產生。 首先使用Pro/E繪製三葉片機械心瓣,然後利用CNC加工出壓克力放大模型及製作原尺寸模型,最後使用體外模擬在脈動流場下觀察心瓣開關行為。 本機械心瓣因葉片變的較輕薄而使心瓣容易開合,並加大樞紐凹槽的孔徑使葉片轉軸圓端能更完整的包覆在凹槽內,加深樞紐凹槽深度讓心瓣更加深入樞紐凹槽內避免葉片脫落,利用樞紐凸塊脊線當成擋板來拘束心瓣葉片的關閉行為。

English Abstract

This research designed trileaflet mechanical heart valve, which opens from the center with a central flow, similar to the function of the human body aortic valve. Compared with the traditional monoleaflet and bileaflet valves, the central orifice of the trileaflet valve increases the effective area, as a result decreasing the pressure drop and the fluid separating. First, use the Pro/E program to design the trileaflet mechanical heart valve, then use CNC to manufacture the enlarge plastics model, and the titanium alloy prototype model. Finally, we observed the opening and closing behaviors in a mock circulate loop. The leaflets of the mechanical heart valve become thinner and easier to open and close. The depths of the fillister are deeper, therefore the leaflet hinges were wrapped up and covered more with the fillister to prevent the leaflets to drop off. When the valve is opened, the leaflet contacts the corresponding face of the protruding hinge. When the valve is closed, the ridge tightly seals with the notch of the leaflet.

Topic Category 工學院 > 機械與機電工程學系碩士班
工程學 > 機械工程
工程學 > 電機工程
Reference
  1. 1.Krucinski S, Vesely I, Dokainish MA, and Campbell G, “Numerical Simulation of Leaflet Flexure in Bioprosthetic Valves Mounted on Rigid and Expansile Stents,”Journal of Biomechanics, vol.20,No8,P.923-943.(1993)。
    連結:
  2. 2.Einav S, Stolero D, Avidor JM, Elad D, and Talbot L,“ Wall Shear Stress Distribution Along the Cusp of a Tri-leaflet Prosthetic Valve,” Journal of Biomechanics,Eng., vol.12,January(1990)。
    連結:
  3. 3.Stein PD, Sabbah MN,“ Measured turbulence and its effect on thrombus formation,”Circulation Research,vol.35,P.608-614.(1974)。
    連結:
  4. 4.Kaster RL,“ Thrombogenicity of prosthetic valves,”The Journal of Thoracic and Cardiovascular Surgery, vol.70,P.539-540.(1975)。
    連結:
  5. 5.Lillehei CW,“ Heart valve replacement with pivoting disc prosthesis. Appraisal of results and description of a new all-carbon model,”Med Instrm, vol.11,P.85-94.(1977)。
    連結:
  6. 6.Hanle DD, Harrison EC, Yoganathan AP, Allen DT, and Corcoran WH,“ In vitro flow dynamics of four prosthetic aortic valves:A comparative analysis,”Journal of Biomechanics, vol.22,No.67,P.597-607.(1989)。
    連結:
  7. 7.Yoganathan AP, Woo YR, and Sung HW,“ Turbulent shear stress measurements in the vicinity of aortic heart valve prostheses,”Journal of Biomechanics, vol.19,P.433-442.(1986)。
    連結:
  8. 13.Leverett LB, Hellums JD, Alfrey CP, Lynch EC,“ Red blood cell damage by shear stress,”Biophysical Journal 12,P.257-273.(1972)。
    連結:
  9. 14.Rooney JA,“ Hemolysis near an ultrasonically pulsating gas bubble,”Science 196,P.869-871.(1989)。
    連結:
  10. 15.Butchart EG, Lewis PA, Grunksemeier GL,“ Low Risk Thrombosis and Serious Embolic Event Despte Low-intensity Anticoagulation:Experience with 1,0004 Medtronic Hall Valves,”Circulation Research, 78(supply):I-66-I-77(1988)。
    連結:
  11. 16.Bellhouse BJ,“ Velocity and pressure distributions in the aortic valve,”J Fluid Mech 37,P.587-600.(1969)。
    連結:
  12. 17.Gillani NV, Swanson WM,“ Time-dependent laminar incompressible flow through a spherical cavity,”J Fluid Mech 78,P.99-127.(1976)。
    連結:
  13. 18.Van Steenhoven AA, Van Dongen MEH,“ Model studies of the closing behaviour of the aortic valve,”J Fluid Mech 90,P.21-32.(1979)。
    連結:
  14. 19.Hart JD, Peters GWM, Schreurs PJG, Baaijens FPT,“ A two-dimensional fluid-structure interaction model of the aortic valve,”J Biomech 33,P.1079-1088.(2000)。
    連結:
  15. 20.W.L.Lim, Y.T.Chew*, T.C.Chew, H.T.Low,“ Steady flow dynamics of prosthetic aortic heart valves:a comparative evaluation with PIV techniques,”Journal of Biomechanics 31,P.411-421.(1998)。
    連結:
  16. 21.Knoch M, Reul H, Kroger R, Rau G,“ Model studies at mechanical aortic heart valve prostheses - Part 1: Steady-state flow fields and pressure loss coefficients,”Journal of Biomechanical Engineer 110,P.334-343.(1988)。
    連結:
  17. 28.Lu Po-Chien, Huang Ren-Hong, Chu Shu-Hsun,“ Tri-leaflet mechanical heart valve,”United States Patent 6,896,700.(2005)。
    連結:
  18. 8.Klepetko W, Moritz A, Mlczoch J et al,“ Leaflet fracture in Edwards-Duromedics bileaflet valves,”The Journal of Thoracic and Cardiovascular Surgery, vol.97,P.90-94.(1989)。
  19. 9.Gross JM, Guo GX, Hwang NHC,“ Venturi pressure cannot cause cavitation in mechanical heart valve prosthetic aortic valves:A comparative analysis,”Trans ASAIO, vol.37,No.3,P.357-358.(1991)。
  20. 10.Hwang NHC,“ Cavitation potential of Pyrolytic carbon heart valve prostheses:A review and current status,” The Journal of Heart Valve Disease, 7(2),P.140-150.(1998)。
  21. 11.Forstrom RJ,“ Membrane strength-the jet fragility test,” Ph D. Thesis, University of Minnesota, U.S.A.(1969)。
  22. 12.Hellums JD, Brown CH,“ Blood cell damage by mechanical forces. In:Hwang NHC, Norman NA, (eds.), Cardiovascular Flow Dynamics and Measurements,”University Park Press,U.S.A.,P.799-823.(1975)。
  23. 22.Po-Chien Lu, Jia-Shing Liu, Ren-Hong Huang, Chi-Wen Lo, Ho-Cheng Lai, Ned H.C. Hwang,“ The Closing Behavior of Mechanical Aortic Heart Valve Prostheses,” Journal of ASAIO, vol.50,No.4,P.294-300.(2004)
  24. 23.Bokros Jack C, Stupka Jonathan C,“ Trileaflet heart valve,”United States Patent 5,843,183.(1998)。
  25. 24.Lapeyre Didier, Steinseifer Ulrich,“ Mechanical heart valve,”United States Patent 6,395,024.(2002)。
  26. 25.Meyer Louis C, Goodenough Samuel H,“ Artificial heart valve,”United States Patent 4,446,577.(1984)。
  27. 26.Milo Simcha,“ Heart valve prostheses,”United States Patent 5,522,886.(1996)。
  28. 27.Bokros Jack C, Stupka Jonathan C,“ Prosthetic trileaflet heart valve,”United States Patent 5,628,791.(1997)。
  29. 29.劉佳興,“主動脈雙葉片人工心瓣脈動流場研究,”博士論文,淡江大學水資源及環境工程研究所(1996)。
  30. 30.顏廉育,“三葉片人工心瓣之設計與體外測試,”碩士論文,淡江大學水資源及環境工程研究所(1997)。
  31. 31.闕永年,“雙葉中心流式人工心瓣膜模型之研究,”碩士論文,台灣大學機械工程學研究所(1994)。