  • 學位論文

利用共振頻率初始值預測牙科植體之癒合時間及可能穩定值:以動物實驗及離體實驗

Predictions of Healing Period & Possible Final RF Values Using Initial RF Values:in Vivo & in Vitro

指導教授 : 李勝揚 黃豪銘
若您是本文的作者,可授權文章由華藝線上圖書館中協助推廣。

摘要


近年來牙科植體的發展已瑧成熟,但卻仍缺乏一套有效裝置可以協助牙醫師評估植體於植入初期,以及在後續的骨癒合過程中的固持狀態。本研究乃分別利用動物實驗與體外模擬實驗,驗證以共振頻率為理論基礎所發展的牙科植體穩固度檢測系統。藉由量測人工植體在不同的穩固度下之共振頻率,以瞭解其未來臨床使用之特性。在為期12週的動物實驗之量測結果顯示,植體植入時的初始頻率與癒合時間及達高原期時的共振頻率上升比例之間,皆存在著近乎線性相關(y = -1583.9x + 12260, R2 = 0.6587及y = -25.949x + 4852.9, R2 = 0.812),且當植入兔子脛骨中的植體之初始共振頻率值為1000~2000 Hz時,可由其量測值增加了90~130 %的量來判斷其已近乎完成癒合;而當植入時之初始共振頻率值為2000~3000 Hz時,或大於3000 Hz時,則可分別由其頻率值增加90 %及50 %來判斷其是否已接近完成癒合。此結果顯示在利用共振頻率法作為牙科植體穩固度參考依據時,可由初始量測值推估完成癒合的時間,並在此時進行第二次量測,再由其增加的比例來判斷骨癒合是否達到完成。且當植體骨整合失敗時,其共振頻率會有下降趨勢。由此可知,共振頻率法不僅可以偵測植體手術骨整合之進行,亦可早期診斷可能的失敗。預期此一新型牙科植體穩固度檢測方法將可成為牙醫師於臨床牙科植體手術時之輔助診斷工具。

並列摘要


In the past thirty years, there were numerous studies published on the stability of dental implant during osseointegration, but there have been still no valid and effective devices yet for dentists to determine the appropriate timing for prosthodontic loading of implants. Recently, implant stability assessment with resonance frequency (RF) has been discussed by numerous studies. However, due to the limit space of oral cavity, the traditional harmonic response method is not suitable for intra-oral practice. Therefore, in the present study, a novel device based on vibration theory developed by our lab was tested for clinical use in vivo and in vitro. The results of RF values show that there are nearly linear relationships between both the initial RF values (IRF) and the healing time and the increasing ratio of RF values (IRRF) at plateau (y = -1583.9x + 12260, R2 = 0.6587 and y = -25.949x + 4852.9, R2 = 0.812). These data demonstrate that the implants with IRF values located at 1000~2000 Hz, 2000~3000 Hz and more than 3000 Hz are stable when their IRRF values reach 90~130 %, 90 % and 50 %, respectively. Therefore, in the clinical practice, we may estimate the healing time of implant by the detected IRF values, and then, make the second detection at the estimated time to judge whether the implant heals from its IRRF value. Furthermore, the RF values might decrease when the osseointegration failed. That is, RF value of implant may help not only to monitor the process of osseointegration of implants, and make the early diagnosis of implant failure, but also estimate the timing of prosthodontic loading. With the optimal modification in the future, it will offer the dentist a useful tool for diagnosis of the dental implant surgery.

參考文獻


25. Huang HM, Cheng KY, Cheng PY, Chen HC, Chen JL, Lian RY, Lee S.Y. The design and verification of stability detecting device for dental implant. Chin Den J. 2003. (in press).
1. Wong FY, Pal S, Saha S. The assessment of in vivo bone condition in humans by impact response measurement. J Biomech, 16(10): 849-856, 1983.
2. Oka H, Yamamoto T, Saratani K, Kawazoe T. Application of mechanical mobility of periodontal tissues to tooth mobility examination. Med & Biol Eng & Comput, 27: 75-81, 1989.
5. Noyes DH, Solt CW. Measurement of mechanical mobility of human incisors with sinusoidal forces. J Biomech, 6: 439-442, 1973.
6. Oka H, Yamamoto T, Saratani K, Kawazoe T. Application of mechanical mobility of periodontal tissues to tooth mobility examination. Med & Biol Eng & Comput, 27: 75-78, 1989.

被引用紀錄


莊瀚伯(2006)。牙科植體術後骨缺損型態之結構分析〔碩士論文,國立中央大學〕。華藝線上圖書館。https://www.airitilibrary.com/Article/Detail?DocID=U0031-0207200917341627
陳梓尉(2007)。共振頻率法於牙根植體缺損位置判別研究〔碩士論文,國立中央大學〕。華藝線上圖書館。https://www.airitilibrary.com/Article/Detail?DocID=U0031-0207200917350513
謝易珊(2012)。結構振型探討牙科植體術後臨床骨缺損〔碩士論文,國立中央大學〕。華藝線上圖書館。https://www.airitilibrary.com/Article/Detail?DocID=U0031-1903201314432261

延伸閱讀