Dose Recalculation and Measurement of MicrosSelectron-HDR 192Ir Source

MicroSelectron-HDR銥192射源的劑量再計算與測量

若您是本文的作者,可授權文章由華藝線上圖書館中協助推廣。

摘要


目的:由於銥192有複雜的γ能譜,所以再計算一個標準包裝的microSelectron-HDR銥192射源之空氣克馬率常數Γδ。用已校正的TLD- 700氟化鋰(LiF)晶片來測量上述射源之空氣克馬。比較劑量的再計算值與測量值。結果亦可應用於新購銥192射源的活度修正。 材料與方法:重新查閱銥192的同位素表之數據後,再計算其空氣克馬率常數Γδ。用銫137射源校正的TLD-700 LiF晶片來測量上述射源之空氣克馬。 結果:對於一個標準包裝的microSelectron-HDR欽192射源而言,考慮0.25 mm厚的不鏽鋼對γ射線所造成的衰滅,空氣克馬率常數Γδ為2.944*10-17Gy m2s-1Bq-1。利用TLD700 LiF晶片來測量10居里(Ci) microSelectron-HDR銥192射源之空氣克馬,所得的結果與再計算的Γδ值相比較。 結論:對於一個新的銥192射源而言,Γδ的再計算值與microSelectron-HDR銥192射源的空氣克馬測量值之間的差異是很有用的。

並列摘要


Purpose: Due to the complex γ- ray spectrum branchings of 192Ir, the air - kerma rate constant Γδor a standard packaged microSelectron-HDR 192Ir source was recalculated. The calibrated TLD-700 LiF chips were used to measure the air-kerma of the above mentioned source. The results of dose recalculation and measurement were compared. The results could be applied in activity correction for a new 192Ir source. Materials and Methods: After reviewing table of isotope data for 192Ir, the air-kerma rate constant Γδwas recalculated. TLD-700 LiF chips calibrated with a standard 137Cs source were applied in the air-kerma measurement of the source mentioned above. Results: For a standard packaged microSelectron-HDR 192Ir source, taking into account attenuation of γrays in 0.25 mm stainless steel, the air-kerma rate constant Γδ was obtained as 2.944*10-17Gy m2s-1Bq-1. Results of air-kerma measurement using TLD-700 LiF chips for a 10 Ci microSelectron-HDR192Ir source was compared with theΓδ recalculation. Conclusion: The difference betweenΓδ recalculation and air-kerma measurement for microSelectron-HDR192Ir is useful in the activity calibration of a new 192Ir source.

參考文獻


Chang, L. L.,Chen, F. D.,Chang, P. S.,Liu, C. C.,Lien, H. L.(1995).Assessment of dose and risk to the body following conventional and spiral computed tomography.Chinese Medical Journal (Taipei).55,283-289.
Ho, A. K.,Sibata, C. H.,deSouza, C. N.,Castiglione, R.,Shin, K. H.(1995).Evaluation of a well-type ionization chamber for calibration of HDR and LDR brachytherapy sources.Medical Dosimetry.20,31-34.
Hubbell, J. H.,Seltzer, S. M.(1996).Tables of X-ray mass attenuation coefficients and mass energy-absorption coefficients from 1 keV to 20 MeV for elements Z.
ICRU(1980).Radiation quantities and units for ionizing radiation.Washington, DC:International Commission on Radiation Units and Measurements.
ICRU(1998).Fundamental quantities and units for ionizing radiation.Bethesda, MD.:International Commission on Radiation Units and Measurements.

延伸閱讀