目的:探討本院兩種透析機在線尿素清除指數(Kt/V)的監測情況,評估其監測Kt/V的可行性,準確性以及在血液透析中的應用價值。方法:利用Gambro透析機以及Dialog透析機在線監測血液透析患者的Kt/V值(online Kt/V);同時抽血檢測血液透析患者前後尿素氮濃度,根據Daugirdas公式計算Kt/V值(spKt/V)作為對照。結果:232例血液透析患者,利用Gambro透析機進行血液透析,測得平均值(online Kt/V)為1.25±1.37;根據Daugirdas公式計算平均值(spKt/V)為1.65±0.31,相關性分析顯示:(r=0.676,P<0.001)。在線測得值(online Kt/V)明顯小於抽血計算值(spKt/V)(P<0.05);而校正後的(adjusted Kt/V)為1.62±0.59,與抽血計算的(spKt/V)值基本接近,相關性分析顯示:(r=0.946,P<0.001)。293例血液透析患者,利用Dialog透析(不提供校正系統)機監測患者的online Kt/V值(1.82±1.27),高於常規抽血方法所得的spKt/V值(1.69±0.37),相關性分析顯示:兩者存在顯著的相關性(r=0.765,P<0.001)。結論:Gambro透析機Diascan監測系統online Kt/V值偏低,校正後Kt/V值卻吻合度好,相關性較好:而Dialog透析機在線檢測值稍高於Kt/V,注意二者的差別,保證患者血液透析的充分性。
Objective. To investigate the feasibility and accuracy of two different dialysis machines monitoring systems in assessing the Kt/V value of hemodialysis adequacy during hemodialysis. Methods. Kt/V values of patients with maintenance hemodialysis were assessed by the monitoring system of Gambro machine and Dialog machine. Meanwhile, blood urea nitrogen (BUN) of pre- and post- dialysis were examined, and the Kt/ V values were calculated according to the Daugirdas formula. The correlation between Kt/ V shown by the monitoring systems and by calculation was studied. Results. The mean value of online Kt/V of 232 hemodialysis patients by Gambro machine (1.25±1.37) was lower than the mean value of spKt/V by calculation (1.65±0.31), and correlation between online Kt/ V and spKt/V was relatively satisfactory (r=0.676, P<0.001). However, there was no significant difference between the spKt/V (1.65±0.31) and adjusted Kt/V (1.62±0.59). The value of Kt/V of 232 hemodialysis patients by Dialog machine (1.82±1.27) was higher than the value of spKt/V by calculation (1.69±0.37), and correlation between Kt/ V by Dialog machine and by calculation was satisfactory (r=0.765, P<0.001). Conclusion. Online Kt/V values by Gambro machine are usually lower than spKt/V calculated by Formula from urea measurements. However, correlation between adjusted Kt/V by Gambro machine and spKt/V by calculation was very satisfactory. Kt/V values by Dialog machine (No adjusting system provided) are usually a little higher than those calculated from urea measurements by Daugirdas formula. This difference has to be considered in application of the results in clinical practice.