  • 學位論文

尿素生化感測器之研究發展

Development of Urea Biosensor

指導教授 : 林孟山

摘要


本實驗利用一氧化二銅(Copper(I) oxide)催化NH3還原的特性,配合Urease (EC 3.5.1.5)可將尿素水解產生NH3,將酵素與一氧化二銅分別修飾在雙玻璃碳電極之上下游,結合流注系統發展低電位氧化模式偵測,並以恆電位儀控制電位及接收訊號,以達發展尿素生化感測器之目的。 此尿素生化感測器之最佳化製備條件是用碳膏與20%一氧化二銅均勻混和後,修飾在下游電極表面,置於40 ℃烘箱30分鐘乾燥後在上游電極滴加0.5 unit Urease (EC 3.5.1.5),靜置於4℃乾燥後再滴1 % 0.5 μl小牛血清蛋白,靜置於4 ℃乾燥後再1 % 0.5μl戊二醛,靜置於4 ℃乾燥即完成電極製備。在最佳化條件下,偵測環境為0.05 M pH 9 碳酸鹽緩衝溶液,偵測電位為200 mV (vs. Ag/AgCl),載體流速為0.5 ml/min,樣品迴路體積為50 μl,進行尿素的量測,所得此感測器分析特性如下:線性範圍為1-10 mM(R=0.99),電流密度為85.736 nA/mM,偵測極限(S/N=3)為85 μM,在精確度方面連續重覆偵測尿素20次操作下,所得到的相對標準差(RSD)為1.8%。且若在雙電極系統的上游電極修飾PbO2進行氧化前處理,可以避免環境中易氧化干擾物如多巴胺、腎上腺素、血清素、組織胺、乙醯苯酚、尿酸、抗壞血酸等。最後將此生化電極與標準方法進行相關性探討,其相關係數為0.995。

並列摘要


The copper(I) oxide based urea biosensor was fabricated in this experiment, and it possessed reduced overvoltage for ammonia determination. The enzyme Urease (EC 3.5.1.5) was drop-coated on the up-stream of dual electrode and subsequently the NH3 was determination through copper(I) oxide modified down-stream electrode. The preparation of this biosensor was as follows, carbon ink and 20% copper(I) oxide had been well mixed and drop-coated on the downsteam electrode ; the urease (0.5 unit) was dropped on the upstream and dried secondly;the 1% 0.5 μl bovine serum albumin drop-coated on the upstream and dried; the 1% 0.5 μl glutaraldehyde dropped on the upstream and dried finally. The optimized conditions in the 0.05M pH 9 carbonate buffer ; applied potential was 0.2 V (vs. Ag/AgCl) ; flow rate was 0.5 ml/min; sample loop was 50 μl. The analytical performances of biosensor equipped with linear range upto 10 mM(R=0.99) and sensitivity is 85.736 nA/mM, detection limit (S/N=3) was 85 μM, and precision is 1.8% by twenty successive measurement. In order to eliminate interference, PbO2 was used to pre-oxidize those easily oxidative compounds, such as ,dopamine, epinephrine, serotonin, histamine , acetaminophen ,uric acid, ascorbic acid. Compared with standard method (sigma640), the correlation coefficient was 0.995.

並列關鍵字

Urea Biosensor Flow injection analysis

參考文獻


Pirvutoiu, L. Valvo, Flow injection analysis of mercury(II) in
Fiber-optic glucose biosensor using enzyme membranes with 2-D
crystalline structure. Biosens. Bioelectron. 11 (1996) 317-325.
changes of immobilized glucose oxidase. Anal. Chim. Acta 319 ( 1996)
3 C. Tran-Minn, D. Valin, Enzyme-bound thermistor as an enthalpimetric

延伸閱讀